Forensic Microscopy
Airborne Particle Sampler Trace Dimensions for Various Air Test Sampling Devices:
This document summarizes the physical properties of ndoor air quality testing & sampling equipment and devices and reviews the methodology for their use.
This is a summary of equipment used for investigating indoor air quality particles (mold, pollen, house dust, insect fragments, dust mites, fiberglass fragments, etc.) found in the the indoor environment.
We include key air test equipment parameters such as particle trace dimensions that are used in calculations of percentage of trace analysis and airborne particle concentration levels. Our page top photo shows the Allergenco Mark II time lapse impaction air sampler in use near an indoor air cleaning machine.
InspectAPedia tolerates no conflicts of interest. We have no relationship with advertisers, products, or services discussed at this website.
- Daniel Friedman, Publisher/Editor/Author - See WHO ARE WE?
IAQ Laboratory SOP(Standard Operating Procedure) Indoor Air Monitoring and Testing Equipment - Particle Sampler Trace calculations recap. This document supports our General Lab Procedures and Particle Trace Count Procedures.
See AIRBORNE PARTICLE ANALYSIS METHODS
Airborne Particle Sampler Trace Dimensions |
|||||
Sampling Device |
Trace length |
Nom. Trace ™ Width |
Flow Rate in LPM |
Pump timer |
Calibrated |
|
|||||
Air-O-Cell™ disposable cassette (Zefon |
14.4mm |
1.1mm |
15 |
2-10 min. typical |
See indiv. Pump device d50=2.6u |
Allergenco D™ disposable cassette (EMS) |
14.0mm nominal |
1.1 mm nominal |
15 lpm recommended, (operates at 5-20 lpm) |
1-10 min. typical |
d50=1.7u |
Cyclex-D™ disposable cassette |
Round media |
<20 lpm |
2-10 min typical |
d50=0.8u |
|
Micro 5™ disposable cassette |
Round media |
5 lpm |
5 min. typical |
1" tube if using Gilair-5 pump d50=0.8u |
|
25mm MCE filter cassette |
25mm MCE filter |
Filter area |
Various |
0.45u (F25045) |
|
37mm MCE filter cassette |
37mm MCE Filter |
Filter area |
various |
0.8u (F37080) |
|
Bi-Air™ disposable cassette |
Dual trace 25mm MCE filter |
Trace area |
3.5 lpm, ™ ™ |
10 min grab, 2-4-hr. TWA sample |
|
Adhesive tape-clear ™ |
Surface particle sampling mold or settled dust/debris Scotch™ Gloss Finish Multi Task Tape 34-8507-5365-7 |
||||
Sterile swab |
Culture collection |
||||
...
IAQ Air Sampling Pumps and Particle Collection Equipment |
|||||
Allergenco Mk3™ time lapse slit-impaction sampler |
14.5mm, up to 12 traces/slide |
1.1mm |
15 nom. lpm |
1min to 24-hrs ™ multi-trace |
5/17/01
|
Burkard PAS#1 |
14.0mm |
2.0mm |
10 lpm |
0-9 min |
Factory d50=2.52u |
Burkard PAS#2 |
14.0mm |
2.0mm |
10 lpm |
0-9min |
Factory |
Intermatic (used with EMS Megalite below) |
NA |
NA |
NA |
1 second to |
Factory |
EMS Megalite Pump w/IAQ Options Ser #7038, Mod. 6025SE-V |
NA |
NA |
3-30 lpm |
Separate duplex timed receptacle |
Factory, pump-mounted rotameter, Lab bubble tube 10/10/05 0 lab< |
Gil-Air-5-RC pump |
NA |
NA |
<1-5 adjustable> |
1 min to 8-hrs |
1/15/04 Lab, Zefon bubble tube (ZBT) + (OD) in field at each use |
PurpleAir PM II Airborne Particle Counter |
PMS5003 laser particle counters also measures humidity & temperature |
Raw PM 0.3um/m3 |
Continuous Real Time PM level alues are averaged every 20 seconds |
Live link AIRBORNE PARTICLE COUNT LIVE DATA SAN MIGUEL de ALLENDE See details at |
|
RadioShack Vacuum Pump |
NA |
NA |
10 lpm |
Manual on-off |
Lab 10/10/2003 ™ -9/5/2005 (ZBT) in lab & (ZR) in field at each use |
Notes: particle collection efficiency d50 for other common devices: (J. Tucker)
AGI-30 liquid impinger ™ 0.31; SAS Slit-to-Agar™0.67; Cyclex impactor 1.80; RCS radial impactor 4.00, Andersen N6 hole impactor 6.61, Rotorod 10.0.
See AIR QUALITY STUDY San Miguel de Allende for a description of use of the PurpleAir PM II - a laser counter device for continuous air sampling and counting of airborne particulates
...
Gas Detection / Sampling Test Equipment Choices |
|||||
Sensidyne Gastec pump (gas detection) |
NA |
NA |
50 or 100 cc/ ™ /full stroke |
Manual stroke, manual count indicator |
Leak test 11/16/04 8/5/2005 |
Draeger accuro pump (gas detection) ARUK-FO13 #1-0893171516K09 exp F013 Serial 6400000 |
NA |
NA |
100 cc/stroke |
Manual, stroke completion indicator, automatic stroke counter |
Leak test 11/16/2004 9/5/2005 |
RTCA Pump#1 Rotary vane vacuum |
NA |
NA |
14.28 |
11.27 min |
Lab 6/11/04 ™ total 160.94L |
RTCA Pump#2 |
NA |
NA |
Tbd |
Tbd |
Tbd |
...
|
|||||
Tramex |
electronic moisture encounter |
||||
Delmhorst |
twin-point electronic resistance moisture meter |
||||
Exergen ™ |
Infrared scanner ™ surface temperature |
||||
OD = flow indicator on device (subject to lab calibration)
ZR = Zefon rotameter, for field use, verified using ZBT in lab
ZBT = Zefon bubble tube (precise), multiple timed events to measure actual LPM
Factory = original calibration provided by factory or independent consultant to factory.
...
Continue reading at AIRBORNE PARTICLE ANALYSIS METHODS or select a topic from the closely-related articles below, or see the complete ARTICLE INDEX.
Or see these
LAB & FIELD IAQ EQUIPMENT SOP at InspectApedia.com - online encyclopedia of building & environmental inspection, testing, diagnosis, repair, & problem prevention advice.
Or see this
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Or see
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Or see
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Or see
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Questions & answers or comments about the properties of test equipment used in indoor air quality investigations for particle hazards such as asbestos, dist mites, dust mite fecals, fiberglass, house dust, insect fragments, mold, pollen, road dust, and other indoor dust and particle components. .
Search the InspectApedia website
Need to make a private comment? Email editor@inspectApedia.com. We welcome email with your website content critique, corrections, suggestions or questions. Email is monitored daily but our workload may prevent an immediate reply. For immediate results, we hope you'll first try asking your question using the search box just above.
In addition to any citations in the article above, a full list is available on request.