EMF Measurement Sensitivity & Accuracy Problems & Solutions
EMF measurement instrument sensitivity & accuracy: this article explains the sensitivity and accuracy of EMF or ELF measuring instruments used for performing electromagnetic field (EMF) or electro-magnetic radiation EMR measurements to measure EMF exposure levels in gauss or milligauss.
We discusses sources of error and variation in EMF measurements and we review and make suggestions for using several low-cost EMF measurement devices to determine the instantaneous electromagnetic field exposure.
InspectAPedia tolerates no conflicts of interest. We have no relationship with advertisers, products, or services discussed at this website.
- Daniel Friedman, Publisher/Editor/Author - See WHO ARE WE?
Because RF and EMF measurement tools need to be properly chosen to measure the particular type and frequency of RF or EMF signal that is of interest, be sure to also
see EMF RF FIELD & FREQUENCY DEFINITIONS
for a simple explanation of different types of radio frequency (RF) and electromagnetic frequency (EMF) types and where they are found.
Most hand-held low-cost EMF measuring instruments (see our list at Evaluation of Low-Cost EMF Instruments) can make measurements down to 1 milligauss.
Because it is common for us to find background EMF levels of 1-3 milligauss in residential neighborhoods where local electrical power distribution lines and transformers are present, and because some studies examined EMF field strengths down to this low level, a do-it-yourself EMF measurement project should probably avoid using instruments that lack that level of sensitivity.
But as we explain next, an accurate, sensitive EMF instrument is insufficient for making accurate and repeatable EMF measurements.
To do otherwise would make it impossible to compare conditions at one property with those at another, even if both properties are equidistant from the same power line and even if both measurements were made at the same moment.
Some EMF measuring instruments, typically called "tri-field" or tri-position instruments, will give a consistent EMF strength reading in a given location regardless of how you are holding the instrument or where you are pointing it - providing the field strength has not changed between measurements.
We recommend this type of instrument for home use by "do it yourself" emf measurers.
See EMF Polarity & Shape Affect Measurements for further explanation.
Other EMF measuring instruments such as our Safeco (Photo at left) are very position-sensitive. The spreadsheets we provide allow for collecting EMF measurements on three axes when using an instrument such as this one.
The disadvantage of the instrument is the requirement to make multiple measurements at each location during an EMF survey.
But the advantage of the instrument is that it allows us to see the shape, source, and even orientation of an electromagnetic field at a given location.
For Position-sensitive EMF measurement instruments, three readings are necessary.
To compute the actual point measurement, each of these numbers, once converted to mG, must be squared, the three squares added, and the square root taken of the sum. This is because the measurement scale is not linear, so a direct raw average would be incorrect. In the EXCEL worksheet which we provide
at EMF MEASUREMENT WORKSHEET you'll see that provision is made for recording raw data points as well as the individual mG readings.
See EMF Polarity & Shape Affect Measurements (just below) for further explanation, and
see EMF MEASUREMENT WORKSHEETfor the proper calculation method used to combine multiple position-sensitive EMF measurements into a single field strength number for a location.
Readers should also see Comparing Gauss versus Milligauss Field Strength Measurements where we explain the greater accuracy of low-level EMF readings when using the milligauss scale on an ELF EMF meter.
Electromagnetic fields are created around power transmission lines by the passage of high levels of current through the transmission line wires themselves. A power-line generated
electromagnetic field has polarity and shape, roughly
spherical around a power line.
The problem of the electromagnetic field having polarity and shape means that some early or low-cost EMF measuring devices will give widely varying field strength measurements depending simply on the physical orientation of the device when the measurement is made - that is, what direction you point the instrument affects its reading.
But don't think that pointing directly towards the power line wires overhead gives the maximum reading.
It may not, due to field polarity. More costly EMF meters have multiple sensors to overcome this defect.
We discuss the problem of electromagnetic field shape, polarity, and instrument testing sensitivity to instrument orientation or position
at Sensitivity & Accuracy of EMF Measuring Instruments.
Instead of contacting us with a request to perform EMF Electromagnetic or RF Radio Frequency Field Strength measurements, in most cases it is more economical and convenient for a property owner to purchase their own instrument, making measurements under varying conditions. In this series of articles we describe how to make measurements using a consistent approach and using good documentation.
See EMF MEASUREMENT STEP BY STEP for details of exactly how to collect EMF measurement data.
Following good procedure and using instruments properly are two steps towards making accurate, repeatable EMF measurements.
But because the signal transmission for RF sources such as radio, TV, or cell towers, the load on a power transmission line is not under control of an individual property owner, and because the EMF strength varies as the power transmission line load varies, it is important to have an idea of that condition as well when attempting to characterize EMF exposure at a specific location.
In contrast, EMF measurements are quite accurate and repeatable at other EMF sources such as close to electrical appliances and service entry cables.
...
...
Continue reading at EMF MEASUREMENT INSTRUMENT USE TIPS or select a topic from the closely-related articles below, or see the complete ARTICLE INDEX.
Or see these
EMF MEASUREMENT INSTRUMENT ACCURACY at InspectApedia.com - online encyclopedia of building & environmental inspection, testing, diagnosis, repair, & problem prevention advice.
Or see this
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Questions & answers or comments about how to hold, position, & use EMF field strength measurement instruments to account for variations in sensitivity and accuracy of Instruments for EMF power line and other source surveys.
Search the InspectApedia website
Need to make a private comment? Email editor@inspectApedia.com. We welcome email with your website content critique, corrections, suggestions or questions. Email is monitored daily but our workload may prevent an immediate reply. For immediate results, we hope you'll first try asking your question using the search box just above.
In addition to any citations in the article above, a full list is available on request.