Test for & Remove Mold Contamination in / on ArtworksMold contamination on or paintings or other artworks.
This article describes using transmitted & reflected light microscopy and polarized light microscopy to test for and report on the appearance of mold contamination on works of art (typically in the museum or in-building environment) and a forensic microscopy approach to detecting mold contamination on artworks in order to assist art conservators in both removing mold contamination and in preventing it on these materials.
We include a discussion of distinguishing among mold and other painting contaminants, stains, including extractive bleeding stains, their causes and prevention. Other sources of biodeterioration of works of art such as cave paintings, and algal growth on outdoor materials are discussed elsewhere.
We link to related articles on paintings or other works of art on or using hardboard such as Masonite™ and soft fiberboard such as Celotex™.
InspectAPedia tolerates no conflicts of interest. We have no relationship with advertisers, products, or services discussed at this website.
- Daniel Friedman, Publisher/Editor/Author - See WHO ARE WE?
- Daniel Friedman, ASHRAE,AIHA, et als. InspectApedia publisher & researcher, the author operated an environmental test and forensic microscopy lab in New York and provided pro-bono mold consulting and advice to several art conservators described herein.
Distinguishing among mold contaminants, other particulate contaminants and other stain sources on or within paintings and other works of art can be an important step in deciding on strategies for both cleaning the artwork and for prevention of future mold, white blooms, effloresence, or other environmental-related damage.
Here we also discuss the distinction of mold contamination from extractive bleeding in art works painted on hardboard products such as Masonite® or Upson Board.
We have assisted several museums & curators with mold diagnosis, cure, & prevention on works of art. The laboratory photographs of mold contaminants in or on artworks are by the author and were made in our forensic laboratory in New York.
The control of indoor humidity where works of art are displayed or stored indoors is naturally the principal step taken to protect these items from mold damage.
Additional steps to protect artworks depends on the materials used and the value of the items.
No valuable artwork should ever be sprayed or coated with a fungicidal sealant without consulting with an expert art conservator, and where an artwork has become mold damaged, amateur cleaning can damage it further - consult with an expert art conservator for advice.
Inoue has tested a varnish coating supplemented with fungicide to attempt to control mold growth on painting surfaces [16]
Krake et als. have evaluated microbiological contamination of artworks studying both the face and the back side of canvases in a museum. [17]
In some cases it may be possible to frame, enclose, or treat the back surface of some paintings without damaging the work itself nor impinging on its artistic or monetary value.
Take care to distinguish between mold growth and other moisture-damage to artworks.
Our photographs above illustrate mold growth on the paper backing of a framed work of art - mold that affects art and artifacts may be hidden on unobserved surfaces but may still be damaging the work.
Our page top photograph illustrates a finding of mold growth (Cladosporium sp. and Penicillium sp.) on the surface of an oil painting. At left, other stains that appeared to be mold needed just a little further examination to distinguish between mold growth and hardboard wood fibers.
In performing analysis for oil paintings displayed in a museum in San Juan, Puerto Rico, working with art conservator Ulrik Runeberg ("Staining and Microbiological Infestation of Acrylic Paintings on Hardboard" [3])
we discovered that "brown spots" that looked like mold contamination were in some cases brown oils or extractive-bleeding from Masonite-type hardboard on which the paintings had been executed.
(Pair of photographs shown above and below.)
Had the artist first sealed the hardboard surface with a suitable sealant, perhaps a lacquer primer, that staining would have been reduced or prevented entirely.
Our photo below shows a higher magnification examination of fibers from the art-work above, as we checked for mold spores or a different cause for discoloration found on the painting obverse side.
Other artworks we have examined suffered severe mold damage, including a wide range of media such as paper-based prints and lithographs or etchings and oil paintings on canvas.
While indeed we often can find mold contamination growing on the surfaces of paintings and other art works, we may also find mold contamination within pigments - possibly from the time of creation of the work - changing the conservator's cleaning and maintenance strategy.
We also often find surface or in-media contaminants that were not mold, though they have been mistaken for it.
These other contaminants include insect fragments, soot particles, road dirt and debris particles, fabric fibers, and even ultra-fine spray paint droplets that, unless examined properly, are mistaken for different contaminants, thus risking misleading the art conservator in her planning for art or artifact restoration or maintenance.
Our photos above and below illustrate an example of potential confusion. By transmitted light (photo above left) the round particle may look like Nigrospora sp. or another fungal spore. But by reflected oblique top lighting it becomes immediately obvious that we are looking at a fine spray paint droplet, in this case of white-pigmented paint.
When appropriate we provide pro bono or fee-paid forensic investigation services to museums and art conservators.
More about the constituents of Masonite type hardboard and an answer to the question of whether or not it contains asbestos is
at HARDBOARD Masonite™-like INGREDIENTS
Question: how to clean mold off of a painting of a New Zealand Tui Bird Prosthemadera novaeseelandiaeI'm an artist living in New Zealand. I have been contacted by the owners of one of my paintings which they purchased five years ago.
A few weeks ago they noticed a white substance they called a mold (which can be seen in the photos attached) on a particular part of the painting.
This particular part of the painting (the bird [a Tui - Ed.]) was done with natural pigments in dispersion mixed with egg yolk. The rest of the painting was done with powered coffee mixed water.
[Click to enlarge any image] Note: this painting excerpt is of a Tui bird common in New Zealand, Prosthemadera novaeseelandiae in the family Meliphagidae.
The painting was given to weeks to dry then it was varnished with two coats of Golden's Archival Varnish MSA w/UVLS (gloss) (spray). I have completed three paintings using these materials and only one has had the issue of mold growth on the egg tempera part.
The owners tell me that the conditions on the room where the painting is kept, are dry and stable. Though [the city where the painting is displayed has experienced it's wettest month in 20 years, so I'm wondering if this may have contributed to the growth of mold.
The main reason I'm contacting you is in regards to how can we remove the mold safely and prevent it from returning. Would re-varnishing the painting after removing the mold help?
Many thanks for your advice. - Phillipa Knight, artist, Gisborne, New Zealand, by private email, 2017/07/31
Phillipa,
While I have some experience with this problem, I'm not an expert painting conservator, any of whom would doubtless have additional and more expert advice.
So before enjoying ourselves with researching and speculating about cleaning egg-tempera paintings by artists of limited means, I'm obligated to say that if you can find a suitable art conservator she could give you the best advice.
You can contact NZCCM, New Zealand Conservators of Cultural Material, Pu Manaaki Kahurangi, Website: http://www.nzccm.org.nz/ where you'll see that Auckland as well as other North Island cities sport art conservators who know about paintings.
The following is based on my own field and lab experience with mold on artwork plus a bit of reading.
In general we need to remove the mold. In the case of a painting we both understand that has to be done with care to avoid damaging the painting.
If you can wipe the mold or debris off of the exterior of the varnished surface then we're in fat city, since we're dealing with a more-superficial cleaning job.
It's possible that the white deposit is a species of mold though I wouldn't rule out a surfacing of a mineral or salt from the original composition of the paint - a "bloom" that's not fungal. Still it will want to be removed.
If you cannot wipe off the mold or debris from the surface of the varnish (say using a dry cotton swab or more aggressive cleaners listed below), then we move on to
Typically a conservator addressing a mold growth under the varnish removes the surface varnish, since otherwise any treatment trying to remove mold below that coating will be ineffective and being more aggressive risks damaging the painting.
The good news is that the Colden Paints Company produce a Golden MSA Solvent that they say can be used to remove the varnish after it is in place on a surface. If you don't have it, here's their contact information
Golden Artist Colors, Inc., 188 Bell Road, New Berlin, NY 13411-9527 USA, 607-847-6154 800-959-6543 Fax: 607-847-6767 help@goldenpaints.com Website: https://www.goldenpaints.com/products/varnish-top-coat/msa-varnish
With the varnish removed you might try very gentle methods to remove the mold, or whatever is contaminating the painting surface, starting with cotton moistened with distilled water first, or perhaps a dilute, mild household detergent.
An aqueous alcohol-water solution made of alcohol and water: specifcially 100% ethanol diluted in distilled water using ethanol/water ratios ranging from 60/40 to 90/10 - the latter, for example, means 90% ethanol/10% distilled water.
Watch out: always start with the mildest-approach to cleaning or conserving any work of art (or building or anything else) so as to avoid unnecessary damage.
Hedley (1980) found a 9% ethanol a good cleaning mix, that's even milder than the alcohol solutions tested by Pruteanu (2014).
The cleaning solution is applied gently with 100% cotton wound on a wooden stick, normally testing it first on a small, unobtrusive area.
Some sources warned that this approach may cause the paint to lose adhesion to the surface.
I suspect that that problem is uncommon and that when it occurs (most sources said it didn't) the problem may be due to either use of too much and too-strong solvent alcohol, leaving it on the surface too long, wetting too deeply, or possibly due to paint that was already poorly-adhered.
Pruteanu et als. describe a detailed study of what formed the most-effective cleaning methods for mold or other contaminants on egg tempera painted surfaces. I'll attach their article as a PDF in the references below. They point out tht the most effective traditional cleaner was described thus:
From the alcoholic solutions tested were chosen only the ones with effective cleaning on degraded varnish layer and deposits of dirt, which was alkalized with three drops of 25% ammonia (E6). - (Pruteanu 2014)
They reported that among the traditional alcohol/water cleaners, their best performer was
... aqueous solution obtained by mixing 6.5 mL distilled water with 2.8 mL ethanol and three drops of ammonia
Those researchers found, for their purposes, even better results with some experimental cleaners made from vegetable products including cabbage juice!
In my opinion it is a serious mistake to try to "kill" the mold: doing so won't work, as to kill all of it you'd damage the painting below with the concentration of disinfectant needed. Other killing approaches using high heat or cold have similar risks for artwork.
I think that the reason you see (what seems to be) mold growth on the egg tempera portions of the painting would be that the particular chemistry of egg-based paint is a bit more mold friendly than the other paints - not that any of them is, without treatment, "mold proof".
Scouring about I found several discussions among artists who use egg tempera in modern paintings, some of whom pointed out that for those who mix oil in with their egg tempera the choice of the particular oil may further invite or retard later mold growth on that work.
Be sure to examine the back of the painting too. I suspect you'll find more, perhaps worse mold there where the local climate and indoor dampness may conspire with the little air circulation at the painting's back to promote mold growth.
The more difficult situation is when mold originated on the painting back (where no one looks) and then grew through the material and paint to the exposed surface. Even where indoor humidity is not extremely high, the lack of air circulation on the back of a painting can make for trouble.
It's also a bit difficult in that the paint itself is organic and supports mold growth. There are (toxic) fungicides that can be added to paint to retard mold growth if you want to try that in the event that you need to touch-up the existing work after cleaning.
Once you're satisfied that the painting is cleaned and repaired, yes you'll want to re-varnish it, making darn sure it's totally dry first since as you probably know, varnishing over moisture will produce a white bloom that's the devil to remove.
Thompson (2012) discusses the importance of varnishing and protecting the surface of paintings. He also warns people using colored earths to build their own earth pigment that you need to wash the sand and clay and organic debris (humus) to make them suitable for pigments.
He mentions, for example, that he found samples of raw sienna that contained enough organic matter to cause them to produce mold growth when used in tempera, also to swell on the panel and ulltimately to “come away”. (Thompson 2005 p. 75)
It’s also important that the painting be completely dry before varnishing. If you varnish over tempera that’s not fully dry a later white bloom under the varnish is likely.
Critical for avoiding mold growth in the future is the correction of the original cause of its appearance.
Varnish on the front of the painting will, properly applied in dry dust-free conditions, give it some protection against occasional surges of humidity.
It's important to look also at the back of any painting suffering from mold or moisture-related damage.
When you inspect the back of the work, if you see mold it needs to be cleaned, though it is not necessary to completely remove all visual traces of the mold. Once you've cleaned the surface of loose actual growing mold structures, the stains that remain, if kept dry, shouldn't give more trouble, and as I suggested above, removing them involves additional risks.
You might, however, improve the mold-resistance of the painting by coating its back surface with a similar sealant, perhaps the same one used on its front. If on inspection you see a mold reservoir on that side that would support this step (after cleaning as best you can).
I know the New Zealand area where your paintin is located. Your weather is not just wetter, you're right on the water, where there's plenty of additional humidity as well as warmer temperatures than down the South Island in Christchurch or Dunedin. (Cooler temperatures also retard many mold genera/species).
The owners may think their home is so clean and dry that their environment isn't a factor, but with respect, they are not quite correct. Airborne mold spores are everywhere, all the time. When mold growth appears on a material it's because conditions were friendly for its growth and development. The chief thing we can control to avoid indoor mold on any surface is the level of moisture in the environment.
My friend Ulrik Runeberg's mold conservation work that we discuss earlier
at ARTWORK MOLD CONTAMINATION worked as a conservator in Puerto Rico on paintings on Masonite - where humidity was causing tannin bleed-through problems, not just mold.
Museums in warm humid climates or in the tropics have a challenge keeping their indoor humidity down. Toledo (2007) reported on another approach to conditioning the whole building: enclosing paintings in glass boxes whose interior was climate-controlled: suitable for museums in the tropics but not for most private collectors buying Kiwi bird paintings.
It's not likely that we can control the environment where your painting will reside, though if they're amenable you could ask your painting-owners to look at their indoor relative humidity and compare it
I'm sure you, can handle this mold on art problem since as we have often been told about solving just about any problem, a Kiwi could do it.
We welcome your further questions, comments, suggestions on this topic and I look forward to seeing photos of your work on the white-mold-suspect bird.
Photos above and below: Tui birds photographed by the website publisher, D. Friedman in 2014.
Photo: An educated Tui singing in the trees of the University of Waikato, Hamilton, New Zeland (DJF 2014).
The control of indoor humidity where works of art are displayed or stored indoors is naturally the principal step taken to protect these items from mold damage.
Watch out: While some experts recommend that paintings and murals must be stored at a relative humidity of 60-70%,[38] our opinion is that those humidity levels are considerably too high to be safe because
...
Continue reading at BOOK / DOCUMENT MOLD & WATER DAMAGE RESTORATION or select a topic from the closely-related articles below, or see the complete ARTICLE INDEX.
Or see these
ARTWORK MOLD CONTAMINATION at InspectApedia.com - online encyclopedia of building & environmental inspection, testing, diagnosis, repair, & problem prevention advice.
Or see this
Or use the SEARCH BOX found below to Ask a Question or Search InspectApedia
Search the InspectApedia website
Need to make a private comment? Email editor@inspectApedia.com. We welcome email with your website content critique, corrections, suggestions or questions. Email is monitored daily but our workload may prevent an immediate reply. For immediate results, we hope you'll first try asking your question using the search box just above.
In addition to any citations in the article above, a full list is available on request.